<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>17004</ErrorNumber>
<ErrorFamily>Exposure Head</ErrorFamily>
<ErrorMessage>
HEAD: Datalink in programming mode, command not valid
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>Thermal</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
The "head program" command has been issued to prepare the head for programming.  Head commands
other than programming commands are not valid. When finished programming, use the "head end"
command.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Record error message and contact Creo Service
</text></DescriptionText>
</OperatorAction>
<Cause>
Non-programming command was issued when head is in programming mode
</Cause>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:33:32 2008 PST
</Updated>
<Links>
</Links>
</Error>